9 Troubleshooting - maintenance

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

78

Fault

Cause

Remedy

Suppressor module /
Sample preparation
module
do not switch to
next position

•

Connection to suppressor
module / sample preparation
module interrupted

•

Suppressor module/Sample
preparation module dry

•

Actuator rotor temporarily
blocked






•

Actuator dirty

•

Actuator has mechanical fault

•

Check cable connection from suppressor
module / sample preparation module to
833 IC Liquid Handling Unit and from 833
IC Liquid Handling Unit to 830 IC Interface
/ 819 IC Detector.

•

Rinse all 3 suppressor units with solution
for at least 10 min.

•

Switch instrument off and on again.
Activate suppressor with <STEP>. Check
that actuator rotor is in correct position
(see Section 9.2.6.3).
Repeat process 2-3 times, before each
switch to next position with <STEP> wait
at least 10 s.
If the suppressor is blocked again then it
must be cleaned or replaced (see below).

•

Clean actuator (see Section 9.2.6.2).

•

Replace actuator (see Section 9.2.6.3).

Unsatisfactory yield for
dialysis

•

Transfer time not determined
correctly

•

Dialysis time too short (e.g. at
too low concentration)

•

Dry or damaged dialysis
membrane

•

Deposits on dialysis
membrane, recognizable by
discoloration at sample inlet

•

Unsuitable dialysis membrane

•

Check optimal transfer time (see Section
7.6.2
).

•

Check optimal dialysis time (see Section
7.6.3
).

•

Replace dialysis membrane
(see Section 9.2.7.1).

•

Replace dialysis membrane
(see Section 9.2.7.1).

•

Use different dialysis membrane.