9.1 Faults and their remedies

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

77

Fault

Cause

Remedy

Inadequate or no supply
at all from pump

Dialysis cell

•

Dialysis cell inlets blocked

•

Sample tubing blocked by too
large particles

•

Bacterial growth in sample
channel and dialysis cell

•

Narrowed PTFE tubing by
compression fittings tightened
too much

Filtration cell

•

Filter membrane blocked,
filtration cell inlets blocked

•

Narrowed PTFE tubing by
compression fittings tightened
too much

•

Replace dialysis membrane and clean
the cell (see Section 9.2.7.1).

•

Replace blocked tubing, always
centrifuge sample (see Section 9.2.4).

•

Replace tubing16, 57 und 58 for sample
channel regularly, replace dialysis
membrane

•

Use capillary cutter to cut off narrowed
tubing ends or replace tubing.


•

Replace filter membrane and clean the
cell (see Section 9.2.8.1).

•

Use capillary cutter to cut off narrowed
tubing ends or replace tubing.

Leaks

•

Tubing nipple leaks

•

Tighten up or replace tubing nipples.

Air bubbles in pump
circulation system

•

Aspiration tubing not immersed

•

Tubing nipple leaks

Dialysis cell

•

Acceptor not properly
degassed

•

Standing air bubbles

•

Completely immerse aspiration tubing.

•

Tighten up or replace tubing nipples.

Dialysis cell

•

Degas acceptor solution again.

•

Small air bubbles do not interfere. For
larger air bubbles rinse the sample
channel with degassed acceptor solution
(possibly open outlet briefly).

Reduced phosphate
sensitivity and/or large
baseline increase
(suppressor)

•

Suppressor capacity reduced
by heavy metal load or organic
contaminants

•

Regenerate suppressor (see Section
9.2.6.1
) or replace it (see Section 9.2.6.3).

Suppressor module /
Sample preparation
module
switches to next
position, but does not
work properly

•

Suppressor module/Sample
preparation module dry

•

Actuator rotor temporarily in
wrong position






•

Actuator step not set correctly

•

Rinse all 3 actuator units with solution for
at least 10 min.

•

Switch instrument off and on again.
Activate suppressor with <STEP>. Check
that actuator rotor is in correct position
(see Section 9.2.6.3).
Repeat process 2-3 times, before each
switch to next position with <STEP> wait
at least 10 s.
If necessary set actuator step again (see
below).

•

Readjust actuator step
(see Section 9.2.6.3).