6.2 Connection to modular IC system for cation separation

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

45

The components are connected as shown below; this also shows the

sample flow scheme:



The inlet and outlet capillaries mounted on the sample preparation

module are connected as follows:

1 Insert sample preparation module

Place sample preparation module 33 inside the chamber on

the floor.

2 Connect sample preparation module

Remove plastic stoppers from the rear panel openings 18 of

the 820 IC Separation Center and lead cable 35 mounted on

sample preparation module 33 back through this opening.

Connect cable 35 to connection 13 “Module” of the 833 IC

Liquid Handling Unit (see Fig. 2).

3 Connect sample feed to sample preparation

module (Sample Preparation connection 1)

Connect inlet capillary 36 (marked with “Sample in”) at

sample preparation module connection 34 (see Fig. 13) to

the 766 Autosampler using the 6.1831.060 PEEK capillary,

two 6.2744.010 compression fittings and a 6.2744.040 PEEK

coupling.

A) Cations of the sample

are continuously
exchanged for H

+

ions

in the ssample
preparation module;
sample loop at valve A
is filled.

Sample

Sample

Column

Column

B) Sample is neutralized

in sample preparation
module; anions are

enriched at valve A.

Fill

A

SP

Module

A

MSM

Inject

A

SP

Module

A

MSM

Regenerant

Eluent

H

2

O

Regenerant

Eluent

H

2

O

Regenerant H

2

O

Regenerant H

2

O

833

SP

818

833

Pump

833

SP

818

833

Pump

Advertising