2.6 High-pressure pump

861 Advanced Compact IC / Instructions for Use 8.861.1033

27

tory filled with either isopropanol or methanol/water). If this is not the

case then the pump must first be rinsed with a solvent which is misci-

ble with both the previous eluent and the following eluent (e.g. ace-

tone).

1 Prepare eluent bottle

•

Prepare, microfilter (0.45

µ

m microfilter) and degas (with N

2

,

He, or vacuum) the suitable eluent for the required applica-

tion and separating column.

•

Fill eluent into eluent vessel 67 (clear glass, 2 L).

•

Place eluent bottle 67 at the front in bottle holder 6 on the 861

Advanced Compact IC (see Figure 1).

2 Install bottle attachment

•

Firmly screw threaded stopper 65 (6.1446.040; part of

6.1602.160) into the smaller threaded opening (M6) of bottle

attachment 66 (6.1602.105; part of 6.1602.160).

•

Firmly screw aspirating filter 68 onto aspirating tubing 63.

•

Pull the other end of aspirating tubing 63 through the larger

threaded opening (M8) of bottle attachment 66 from below.

•

Push O-ring (E.301.0021; part of 6.1602.160) over the free

end of aspirating tubing 63 and move it towards bottle at-

tachment 66.

•

Push tubing nipple 64 (4.420.4300; part of 6.1602.160) over

the free end of aspirating tubing 63, move it as far as requi-

red towards bottle attachment 66 and screw it loosely in the

larger opening of bottle attachment 66.

•

Insert aspirating tubing 63 with screwed-on aspiration filter 68

into eluent bottle 67 and screw bottle attachment 66 onto

eluent bottle 67.

•

Pull aspirating tubing 63 so far through the opening of tubing

nipple 64 that aspirating filter 68 is touching the bottom of

eluent bottle 67.

•

Fix aspirating tubing 63 in place by screwing shut tubing

nipple 64.

3 Mount CO

2

absorber tube

•

First place a piece of cotton wool 70 followed by CO

2

ab-

sorber 69 (e.g. Merck soda lime pellets with indicator, no.

6839.1000) in the large opening of absorber tube 72 and then

close it with the plastic lid.

•

Fasten absorber tube 72 to bottle attachment 66 with the aid

of SGJ clip .

4 Connect aspirating tubing to pump

•

Insert the free end of aspirating tubing 63 into one of the ope-

nings 8 in the interior of the 861 Advanced Compact IC (see

Figure 2).

•

Pull aspirating tubing 63 sufficiently far into the interior of the