Metrohm 861 Advanced Compact IC User Manual

Page 61

Advertising
background image

2.11 Connection of external devices

861 Advanced Compact IC / Instructions for Use 8.861.1033

51

On Version 2.861.0040 the 853 CO

2

Suppressor is already

connected to Remote interface

21

. Connect the 6.2141.190

Cable to the rear of the Remote connector.

2 Tubing connection 813 – Injection valve

•

Screw off aspirating tubing 29 from injection valve 28 and

remove from feedthrough 4.

•

Cut the PEEK capillary tubing mounted on the 813 Compact

Autosampler to the required length (so that it can be connec-

ted to injection valve 28 of the 861).

•

Pull the free end of the PEEK capillary tubing through open-

ing 8 of the 861 Advanced Compact IC and use a 6.2744.010

PEEK Compression fitting to screw it onto injection valve 28.

3 Tubing connection Injection valve– Waste

•

Insert 6.2744.020 Coupling (from the 861 accessories) into

connection 3 of the 861 Advanced Compact IC.

•

Screw aspirating tubing 29 into the 6.2744.020 Coupling and

insert into the waste container.

4 Setting a method on the 813

•

Use the 813 keyboard on the 813 Compact Autosampler to

set the required method (see 813 Instructions for Use).

5 Creating a time program for the 861

•

In the main window of the «IC Net» software select

File / Open /

System

. Open the system for the 861 Advanced Compact IC

that you have created in Section 2.5.4.

•

Open the window with the system settings of the 861 (with a

double-click on the 861 icon or via the right-hand mouse key

open

).

•

Open the

Program

tab.

•

Create a time program:

The Remote command sends a signal to the 813 via Remo-

te line 3. The function of this signal depends on the method

Advertising