4.3 Faults and malfunctions

861 Advanced Compact IC / Instructions for Use 8.861.1033

83

Malfunction

Cause

Rectification

Extreme peak
broadening,
splitting (double
peaks)

•

Dead volume at the co-
lumn ends

•

Check the connections
(use PEEK capillaries with
0.25 mm i.d. between in-
jection valve and detector
block)

•

Turn separation column (if
allowed), or replace it

No feed of re-
generation or
rinsing solution
for suppressor

•

Contact pressure too
low

•

Leak in system

•

Defective pump tubing

•

Contamination of the fil-
ter in the 6.2821.120 Fil-
ter unit PEEK

•

Counterpressure at sup-
pressor module «MSM
II» too high

•

Adjust contact pressure
(see Section 2.10.2)

•

Check connections and
make leakproof

•

Replace pump tubing (see
Section 4.2.10)

•

Replace the 6.2821.130
Filter (see Section 2.3.6)

•

Clean or replace suppres-
sor (see Section
4.2.7

…

4.2.9)