2 Installation
861 Advanced Compact IC / Instructions for Use 8.861.1033
46
•
Select
Connect to workplace
in the
Control
menu in this window.
2 Open control window
•
Double-click the 861 icon in the system window. The control
window for manual control of the 861 Advanced Compact IC
appears, which indicates conductivity, pressure and current
system parameters.
Make sure that the set flow does not exceed the maximum
permissible flow for the particular column (see column data
sheet).
3 Start system
•
Make sure that the aspirating tubing 63 for the high-pressure
pump is immersed in the eluent.
•
Select
Startup hardware (Measure baseline)
in the
Control
menu
of the system window. The high-pressure pump, peristaltic
pump and, if present and activated under
Configuration
(see