Alarm stops
The
Alarm stops
tab of the hardware settings window defines the events
for which the instrument is stopped immediately. At an alarm stop, high-
pressure pump and peristaltic pump are stopped immediately, the run-
ning determination and the active sample queue are also stopped.
Event
Events for alarm stop:
Pump stop
Pump stopped because pressure limits are ex-
ceeded.
Hardware error
Hardware error detected at the 861 Advanced
Compact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Leak
Leak detector has detected solvent in the in-
strument’s interior. This information is also
stored in the instrument itself, so that it is
stopped automatically even without connection
to the PC.
Power on
Temporary power failure at the 861 Advanced
Compact IC hardware.