2.10 Putting into operation

861 Advanced Compact IC / Instructions for Use 8.861.1033

47

Section 3.3.1), the 853 will start up. At the same time a chro-

matogram window opens in which the baseline is shown con-

tinuously.

4 Set contact pressure for pump tubings

•

Press contact pressure lever 48 on both tubing cartridges 47

upwards until regeneration and rinsing solution just start to be

drawn in.

•

Then press contact pressure lever 48 upward by 1 more click

in order to achieve an optimal contact pressure.

5 Check for leaks

•

Check all capillaries and their connections between the high-

pressure pump and the detector block and between the peri-

staltic pump and the suppressor module «MSM II» for escap-

ing liquid. If eluent escapes anywhere then the appropriate

compression fitting must be tightened further or changed.

6 Condition system

•

Rinse the system with eluent until the desired stability of the

baseline is reached (normally 30

…

60 min; if the eluent is

changed, the establishment of the ion exchanger equilibrium

on the separating column can take longer). During that time,

step the suppressor at least 3 times (approx. every 10...20

minutes), so that after this time the suppressor module is suf-

ficiently conditioned too.

•

The instrument is now ready for sample determinations using

the selected system.

Pump tubings are consumable material with a lifetime which depends

on the contact pressure. This is why the tubing cartridges should be

raised completely by loosening snap-action lever on the right-hand si-

de if the pump is to remain switched off for a considerable length of

time (the set contact pressure remains unchanged).