2.8 Suppressor module «MSM II»
861 Advanced Compact IC / Instructions for Use 8.861.1033
41
module «MSM II» 46 from below through one of the openings
8 out of the inner compartment of the 861 Advanced Com-
pact IC.
•
Lead outlet capillary 87 to a sufficiently large waste container
and fix it in place.
Figure 18:
Connections at suppressor module
«MSM
II»
5 Inlet capillary for H
2
O
•
Attach inlet capillary 84 marked with "H
2
O" of suppressor mo-
dule «MSM II» 46 using a 6.2744.010 Compression fitting to
the PEEK Coupling 81 connected to the front pump tubing
79.
6 Outlet capillary for H
2
O
•
Pull outlet capillary 86 marked with "Waste" of the suppressor
module «MSM II» 46 from below through one of the openings
8 out of the inner compartment of the 861 Advanced Com-
pact IC.
•
Lead outlet capillary 86 to a sufficiently large waste container
and fix it in place.
7 Fasten capillaries to the side walls
•
If necessary, the two aspirating tubings 75 and 76 can be fi-
xed in the required position in the interior with the help of a
Y.107.0150 self-adhesive strap.
•
If necessary, the two outlet capillaries 86 and 87 can be fixed
in the required position in the interior with the help of a
Y.107.0150 self-adhesive strap.
82 Suppressor inlet capillary
for eluent
83 Suppressor outlet capil-
lary for eluent
84 Suppressor inlet capillary
for H
2
O
85 Suppressor inlet capillary
for H
2
SO
4
86 Suppressor outlet capil-
lary for H
2
O
87 Suppressor outlet capil-
lary for H
2
SO
4
1
3
2
86
84
83
82
85
87
Waste
Eluent
H
2
O
Detector
Waste
H
2
SO
4