10

Technical

Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

To ensure proper communication between electrophoresis station and software, always power on
the electrophoresis station before launching the software.

2.2.3 Computer (PC)
Depending on the system configuration ordered, your shipment may or may not include a PC. Connect
the PC directly to a power outlet and power it on.

If using your own PC, ensure it meets the following specifications:

n

Operating system Windows XP (service pack 3) or Windows Vista (service pack 1)

n

Processor

3 GHz or higher

n

Memory

1 GB RAM or greater

n

Free disk space

80 GB or greater

2.2.4 Software
Insert the Experion software CD-ROM and follow the prompts to install Experion software on the
PC. The installation program places an application icon on your desktop and creates a folder named
Experion under Programs in the Windows Start menu.

After installation, reboot your computer and confirm that Experion software has been installed. Do not,
however, launch the software.

Fig . 2 .4 . Removing the chip from the Experion vortex station .

2.2.5 Electrophoresis Station
The electrophoresis station is equipped with a universal power supply to operate at 100–240 VAC,
50–60 Hz. Connect the electrophoresis station to an outlet and to your PC:

1. Place the electrophoresis station near the PC, on a sturdy surface and away from direct light and

all other potential sources of extreme heat. Place it where it will not be subjected to vibrations or
other movement.

2. Connect one end of the power cable to the electrophoresis station and the other end to a grounded

AC outlet rated 10 A/100–240 V (Figure 2.5).

3. Connect the electrophoresis station to the PC using the USB2 cable (Figure 2.5).

4. Power on the Experion electrophoresis station by pushing the green button on the front of the station

(Figure 2.6). The green LED above the green button illuminates, indicating the unit is on. Use the
Hardware Wizard to install the driver for the electrophoresis station onto the PC.

Experion Automated Electrophoresis System