4.2 Priming Station

Error

Probable Cause

Recommended Action

LCD screen is blank

There is no power to the priming station

Check the power cord connections and for AC power to the
outlet

Check Seal message

The priming seal is worn, torn, or clogged

Replace the priming seal (see Section 3.2, Priming Station)

The chip is not positioned correctly in the
priming station

Align the arrow on the chip with the arrow on the platform

The gel-stain solution (GS) was added to the
wrong well for priming

The priming well is highlighted on the chip. Refer to the
assay manual for information on how to prime the chip

Nothing happens when the
Start button is pushed

The gel-stain solution (GS) was added to the
wrong well for priming

The priming well is highlighted on the chip. Refer to the
assay manual for information on how to prime the chip

Timer fails to count down
to zero after pushing the
Start button

The priming seal is worn, torn, or clogged

Replace the priming seal (see Section 3.2, Priming Station)

Service message and
wrench icon appear in
the display

Contact your local Bio-Rad office

After priming, a bubble
is present in the
microchannels

Poor application of gel-stain solution (GS);
bubbles that are forced into the microchannels
take the shape of the microchannel and are
elongated, not round

Prime a new chip

After priming, channels
are not filled with gel-
stain solution (GS)

Inappropriate priming pressure and time
settings were used

Use the pressure and time settings indicated on the chip for
the selected assay

GS was added to the wrong well for priming

The priming well is highlighted on the chip. Refer to the
assay manual for information on how to prime the chip

The priming seal is worn, torn, or clogged

Replace the priming seal (see Section 3.2, Priming Station)

Particulates in GS are clogging the
microchannels

Prepare new GS or refilter an older aliquot

After changing the priming seal, or if the display is not working properly, reset the priming station by
unplugging it for a few sec and then reconnecting the power cord.