3 vortex station, 4 software errors, Experion automated electrophoresis system – Bio-Rad Experion DNA Analysis Kits User Manual

Page 26

Advertising
background image

4.3 Vortex Station

Error

Probable Cause

Recommended Action

No movement or sound

There is no power to the vortex station

Check the power cord connections and for AC power to the
outlet

Unplug the power cord and check both fuses

If all of the above are fine, contact Bio-Rad Technical Support

Vortex adapter is damaged
or does not move but motor
makes some noise, or if
solution splashes out of chip
wells after vortexing

Contact Bio-Rad Technical Support

4.4 Software Errors

Error Code

Message

Probable Cause

Recommended Action

50001

Login failed

The username and/or password
is invalid

Log in again using a valid username and
password

50004

Failed to create the data file
folder

Log in as a Windows Administrator and reinstall
Experion software

50005

Failed to create the packet file
folder

Log in as a Windows Administrator and reinstall
Experion software

50006

Cannot start a run because the
electrophoresis station is in use

Wait for a run to complete before starting a
new run

50007

Failed to locate the
electrophoresis station selected

Confirm the electrophoresis station and PC are
connected and try again

50008

Failed to open the assay selected

Restart the computer and try again

Reinstall Experion software

50009

Cannot find the script for the
assay selected

Reboot the computer and run the assay again; if
the problem persists, reinstall Experion software

50011

The selected assay cannot
be run on this electrophoresis
station

The software is in simulation
mode; an assay cannot be run
on a simulation electrophoresis
station

Deselect simulation mode to run the assay.
Select Tools > Options > Advanced and then
deselect Instrument Simulation Mode

50013

Failed to create the data file

Check permissions to write to the data folder

50015

Failed to locate the run selected

The run information appears in
the tree view, but the data file is
missing

Import the data file again by browsing to the data
file location

50016

No assay selected

Select an assay, then start the run

50017

Unexpected error occurred in the
application. Please check the log
file for details

Take the action indicated in the system log file

22

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

Experion Automated Electrophoresis System

Advertising