Error Code

Message

Probable Cause

Recommended Action

50019

The run is already present in the
tree view

An attempt was made to import
a run that is already present

Check the tree view for the imported run

50021

Failed to save the run

Check permissions to write to the data folder

50022

Cannot change the
electrophoresis station
simulation mode because
an electrophoresis station is
currently running

The software is trying to toggle
between normal and simulation
mode, which cannot be done
while the electrophoresis station
is running

Stop the run to switch modes. To switch modes,
select Tools > Options > Advanced and then
select or deselect Instrument Simulation Mode

50024

Select a project before starting
a run

Select a project, then start the run

50033

Failed to connect to the database

See the Security Edition startup guide or
software Help file for requirements for connecting
to the database

50046

Aborting the run

Lid is opened during the run

Do not open the lid during the run; to stop the
run, click Stop Run

50047

Cannot close Experion because
electrophoresis stations are
acquiring data

Wait for the run to complete before exiting

50060

Failed to save protein names

Check permissions to write to the data folder

50063

Cannot start run -
Electrophoresis station lid open

Close the lid and start the run

50064

Cannot start run - no chip
detected

Confirm that the chip is primed and loaded
properly and that it is seated correctly in the
electrophoresis station. Then restart the run

50068

Failed to subscribe to
electrophoresis station control
events

Restart the computer and try again

Disable the firewall

50075

Cannot logoff because the user
is currently running experiments

Wait until the run is complete prior to logging off

50081

Cannot perform software OQ
because configuration files are
missing

Login as a Windows Administrator and reinstall
Experion software

Application needs to be run with
administrative privileges

Right-click the Experion software desktop
icon to launch software, and select Run As
Administrator

Attempt to start assay without
script

A script was not received by the
electrophoresis station

Restart Experion software

BsnReader has stopped working

Restart the computer and power off/on the
electrophoresis station

Cannot communicate with
channel power supplies

Contact Bio-Rad Technical Support