Page 28

Advertising
background image

Error Code

Message

Probable Cause

Recommended Action

Database at <path> could not be
attached

The database file is missing or
corrupted

Restore from backup or recreate by reinstalling
software

Database server is not available
or not found

Restart the computer and try again. If the
problem persists, update Windows via Windows
update and then uninstall Experion software (see
Uninstalling Experion 3.0 XP in the software Help
file). Reinstall Experion software

Did not receive a proper
response to a command

The PC and electrophoresis
station are not synchronized

Exit the application and reboot the PC

Power off/on the electrophoresis station
and restart the software. Power on the
electrophoresis station before launching
Experion software

If the problem persists and data are affected,
contact Bio-Rad Technical Support

Failed to add user < LoginName>
to Windows ‘Experion’ Group

Insufficient permissions

See the Security Mode section of the software
Help file

Failed to add Windows ‘Experion’
Group

Insufficient permissions

See Launching the application for the first time
for Vista or XP in the software Help file

Failed to add Windows ‘Experion’
group access to Experion
Database

Insufficient permissions

See the Security Mode section in the software
Help file

Restart the computer and try again. If the
problem persists, update Windows via Windows
update and then uninstall Experion software (see
Uninstalling Experion 3.0 XP in the software Help
file). Reinstall Experion software

Failed to add Windows
‘Experion’ Group - Windows
Logon User must be a Windows
Administrator

Insufficient permissions

See Launching the application for the first time
for VISTA or XP in the software Help file

Failed when checking if Experion
Database is attached.

Insufficient permissions

See the Security Mode section of the software
Help file

Restart the computer and try again

Update Windows via Windows update and then
uninstall and reinstall Experion software (see
Uninstalling Experion 3.0 XP in the software Help
file)

Focus Motors and Limit Switches
failed

Contact Bio-Rad Technical Support

Installation rolls back

Corrupt Setup.exe file

Install the software using a different Setup.exe file
(contact Technical Support for help)

Antivirus software blocking
installation

Disconnect the computer from the Internet
to minimize the chance for virus attack, stop
antivirus software, and try again

Bad conditions (for
example, corrupted registry,
malfunctioning components, etc.)
in the operating system

Ask your IT department to check the integrity of
the operating system

24

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

Experion Automated Electrophoresis System