Analyzing Data Using Cutoffs

47

Reference can be made to standard or sample IDs that have been defined in the
Template setup

• Standards: enter ! followed by concentration
• Unknowns: enter identifier
• Blanks: enter

#

In all Cutoff equations, Microplate Manager 6.0 uses the averaged value of that
group.

Examples of valid expressions:

• U01
• (U01+.1)/2
• (!100-!10)*2

Cutoff Examples for Assays with Controls

Positive & Negative Controls – positive (POS) & negative (NEG) control well(s)
defined on template

• Low Cutoff: NEG
• High Cutoff: POS

% B/Bo – non-specific binding blank (NSB) & maximum or Bo well(s) (MAX)
defined on template

• Low Cutoff: NSB
• High Cutoff: MAX

% Calibrator –blank (#) & calibrator (CALB) well(s) defined on template

• Low Cutoff: #
• High Cutoff: CALB

% Control - non-specific binding blank (NSB) & control well(s) (CNTRL) defined
on template

• Low Cutoff: NSB
• High Cutoff: CNTRL