Setting up the equations – Bio-Rad Microplate Manager Software User Manual
Page 74
Advertising

Appendix
66
Setting up the Equations
1. Click Template View icon, select Equation from popup.
2. Click Equation Editor icon. Choose % Max from Equation Name popup
to view equations to be applied sequentially to data.
@ - # subtracts mean of blank (#) wells
@ - NSB subtracts mean of non specific binding (NSB) wells
@ / MAX divides by mean of maximum binding (MAX) wells
@ * 100 multiplies by 100 to convert to % bound
Advertising