Welch Allyn CP 200 12-Lead Resting Electrocardiograph - User Manual User Manual

Page 102

Advertising
background image

98

Appendix B EMC Guidance and Manufacturer’s Declarations

Welch Allyn CP 200 Electrocardiograph

Table 6. Electromagnetic Immunity

The CP 200 electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the CP 200
electrocardiograph should assure that it is used in such an environment.

Immunity Test

IEC 60601
Test Level

Compliance Level

Electromagnetic Environment - Guidance

Electrostatic discharge
(ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.

Electrical fast transient/
burst

IEC 61000-4-4

±2 kV for power supply
lines

±1 kV for input/output
lines

±2 kV for power supply
lines

Not applicable

Mains power quality should be that of a typical commercial or
hospital environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should be that of a typical commercial or
hospital environment.

Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines.

IEC 61000-4-11

>95% dip in 0.5 cycle

60% dip in 5 cycles

30% dip for 25 cycles

>95% dip in 5 seconds

>95% dip in 0.5 cycle

60% dip in 5 cycles

30% dip for 25 cycles

>95% dip in 5 seconds

Mains power quality should be that of a typical commercial or
hospital environment. If the user of the CP 200
electrocardiograph requires continued operation during power
mains interruptions, it is recommended that the CP 200
electrocardiograph be powered from an uninterruptible power
supply or battery.

Power frequency
(50/60Hz)
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.

Advertising