Chapter 8 Troubleshooting

91

AC interference (even-peaked, regular
voltage superimposed on the waveforms).
May resemble or coincide with muscle-
tremor interference.

Electrodes that are dirty, corroded,
loose, or positioned on a bony area.

Insufficient or dried electrode gel.

Patient or technician touching an
electrode during recording.

Patient touching any metal parts of an
exam table or bed.

Broken lead wire, patient cable, or
power cord.

Electrical devices in the immediate
area, lighting, concealed wiring in
walls or floors.

Improperly grounded electrical outlet.

Incorrect AC filter frequency setting or
AC filter is turned off.

Check all electrode contacts and lead wires.

Verify that the patient is not touching any metal.

Verify that the AC power cable is not touching
the patient lead cable.

Verify that the proper AC filter is selected. See

“Mains Filter”

on page 48.

If interference persists, unplug the
electrocardiograph from AC power and run it on
the battery. If this solves the problem, you’ll
know that the noise was introduced through the
power line.

If interference still persists, the noise may be
caused by other equipment in the room or by
poorly grounded power lines. Try moving to
another room.

Table 2. Lead Quality Problems

(Continued)

Condition

Causes

Actions

Table 3. System Failure Problems

Condition

Causes

Actions

Won't turn on when plugged into AC
power.

Faulty AC power connection.

Blown AC fuses.

No AC power.

Check the AC power source.

Check the AC fuses. See

“Replacing the AC

Fuses”

on page 86.

Won't turn on when unplugged from AC
power.

Battery disconnected or incorrectly
connected.

Battery low, not charging, depleted, or
bad.

Blown battery fuse.

Check battery connections. See

“Replacing the

Battery”

on page 83.

Recharge the battery. See

“Recharging a Fully

Discharged Battery”

on page 82.

Replace battery. See

“Replacing the Battery”

on

page 83.

Replace battery fuse. See

“Replacing the Battery

(DC) Fuse”

on page 85.

Shuts down during printing

Battery low or bad.

Recharge the battery. See

“Recharging a Fully

Discharged Battery”

on page 82.

Replace battery. See

“Replacing the Battery”

on

page 83.

Prints fewer than 10 reports on a full
battery charge.

Degraded battery.

Replace battery. See

“Replacing the Battery”

on

page 83.

Unresponsive for an extended time.

System frozen.

Press the reset button by inserting a small object,
such as a paper clip, into the hole.

The screen is too dark or too light.

The contrast is to high or too low.

Press Ctrl-R. The contrast resets to its default value.

Reset button