Chapter 8 Troubleshooting

Welch Allyn CP 200 Electrocardiograph

Table 4. System Messages (alphabetical order)

System Message

Problems

Actions

“A test is being printed that has not been
saved and will be lost. Continue
shutdown?”

was pressed while printing an

Auto ECG.

To shut down without saving the test, press

.

To cancel the shutdown, press

.

“Audit Trail is too large. Please print and
purge.“

Audit information is approaching its
maximum allotted storage.

Print and purge the audit trail. See

“Working With the

Audit Trail”

on page 78.

“Insufficient battery power to begin.
Please connect AC power and try again.”

Battery low or bad.

Recharge the battery. See

“Recharging a Fully

Discharged Battery”

on page 82.

Replace battery. See

“Replacing the Battery”

on

page 83.

“Insufficient space available”

Not enough space on memory card
or removable USB storage device.

Delete some tests from the card or USB storage
device at a PC.

Use a different card or USB storage device.

“Memory card error”

Problem writing to memory card
or removable USB storage device.

Verify that the write-protect tab is in the
unprotected position.

Reseat the card in its slot.

Use a different card or USB storage device.

“Out Of Paper”

Printer is out of paper.

Printer door is open.

Load paper. See

“Loading the Thermal Chart

Paper”

on page 23.

Close the printer door. See

“Loading the Thermal

Chart Paper”

on page 23.

“Paper Error”

Paper was loaded incorrectly.

Reload the paper. See

“Loading the Thermal Chart

Paper”

on page 23.

“Powering down”

Low battery.

Recharge the battery. See

“Recharging a Fully

Discharged Battery”

on page 82.

“Problem loading the following settings:
<System>
<ECG>
*
*
*
Using default settings.”

Problem loading your settings at startup.
May indicate memory problems.

Contact Technical Support. For phone numbers, see

page ii

.

“Shutdown?”

was pressed while printing a

Rhythm ECG.

To shut down, press

.

To cancel the shutdown, press

.

“Temperature Error”

Printer head temp is too high.

Allow to cool, then try again.

“Test has not been saved. Continue
shutdown?”

was pressed after printing an

Auto ECG.

To shut down without saving the test, press

.

To cancel the shutdown, press

.

“Unable to communicate with
workstation“

Configuration not correct.

Cable and/or wireless radio not
properly connected.

Verify communication settings on the PC.

Check cable and/or wireless radio connections.