54

Chapter 5 Performing ECG Tests

Welch Allyn CP 200 Electrocardiograph

Figure 33. Electrode Placement Locations — Augmented Pediatric

Electrodes

AHA

IEC

Locations

A

V1

red

C1

red

Fourth intercostal space at right sternal border.

B

V2

yellow

C2

yellow

Fourth intercostal space at left sternal border.

C

V4

blue

C4

brown

Fifth intercostal space at left of midclavicular line.

D

V7

orange

C7

black

Posterior axillary line at the same horizontal level as V4.

E

V6

purple

C6

purple

Mid-axillary line on same horizontal level as V4.

F

LA

black

L

yellow

Just above left wrist on inside of arm.

G

LL

red

F

green

Just above left ankle.

H

RL

green

N

black

Just above right ankle.

I

RA

white

R

red

Just above right wrist on inside of arm.

J

V3R

green

C3R

green

Mirror position to standard V3 placement with respect to the mid-
sternal line.

I

A

H

B

J

C

E

D

F

G

To use the augmented pediatric lead set, you
must enable it in the settings. See

“Turning the

Augmented Pediatric Lead Set On and Off”

on

page 49.