Recording a stat auto ecg – Welch Allyn CP 200 12-Lead Resting Electrocardiograph - User Manual User Manual

Page 66

Advertising
background image

62

Chapter 5 Performing ECG Tests

Welch Allyn CP 200 Electrocardiograph

Recording a Stat Auto ECG

A stat Auto ECG is an immediate printout in Auto Report 1 format.

Stat mode bypasses patient data entry, as shown in

Figure 36

on page 56. A temporary

ID number is assigned to the patient to identify stat tests. After printing, you may enter
the patient data by editing the test.

In stat mode, Auto Send and Auto Save are always disabled, even if they are enabled in
the ECG settings. If you want to send or save a stat Auto ECG, you may do so manually
after you print.

To Record a Stat Auto ECG

1.

Press and hold the Auto ECG key

.

The electrocardiograph begins acquiring ECG data. After it has acquired 10 seconds of
quality data, it prints a report.

2.

Go to

Step 7

on page 58.

Continue the procedure as if you have just pressed the Print ECG softkey.