Chapter 7 Maintaining the Electrocardiograph
83
Replacing the Battery
If you have recharged the battery and the electrocardiograph still does not turn on when
unplugged, or if the battery loses its charge quickly, replace the battery as follows.
(For part number, see
1.
Unplug the electrocardiograph from AC power if connected.
2.
Turn the electrocardiograph upside-down.
3.
Unscrew and remove the battery door. See
Figure 50
.
4.
Lift out the battery. See
Figure 51
.
5.
Inspect the fuse. See
•
If the “Z” wire is intact, go to
Step 6
.
•
If the “Z” wire is broken or dark, replace the fuse. See
Figure 52. Battery Fuse Plugged In
Figure 50. Removing the Battery Door
Figure 51. Lifting Out the Battery
“Z” wire