Specifications, A - specifications, Fuses – Welch Allyn CP 200 12-Lead Resting Electrocardiograph - User Manual User Manual

Page 99

Advertising
background image

A

95

Specifications

Item

Specification

Dimensions

Weight

11.6 lb (5.3 kg)

Keyboard type

Elastomer keypad with complete alphanumeric keys
Soft-key menu and dedicated function keys

Paper type

8.25 x 11 inches (21 x 28 cm), Z-fold thermal paper, 200 sheets

Thermal printer (internal)

Computer-controlled dot array, 8 dots/mm

USB printer (external)

Inkjet or laser printer that supports PCL (printer control language)

Thermal paper speeds

10, 25, 50 mm/s

Gain settings:

Auto ECGs
Rhythm ECGs

5, 10, 20 mm/mV, AUTO
5, 10, 20 mm/mV

Report print formats:

Auto ECGs
Rhythm ECGs
Average cycles

3x4, 3x4+ 1R, 3x4 +3R, 6x2, 12x1, 6x2 50 mm/s, 6x2 Ext., 2x6 +1R, 6x2 +1R
3, 6, or 12 leads at a time
3x4 50 mm/s + 3R, 6x2 50 mm/s + 6R

ECG storage

Up to 50 ECGs

Frequency range

0.3 to 150 Hz

Digital sampling rate

> 1,000 samples/second/channel

Pacemaker detection

ANSI/AAMI EC11

Power requirement

Universal AC power supply 100–240 V ~, 50/60 Hz ~, 2.2 A maximum

Fuses

AC
DC (battery fuse)

Time-delay type, 2 amp 250 V rating, Littlefuse 0215002 or equivalent.
Fast-acting type, 10 amp 32 V rating, Bussman ATC-10 or equivalent.

Lead configurations

Standard, Cabrera

15.6 in
(39.7 cm)

6.2 in
(15.7 cm)

16.2 in
(41.1 cm)