FIG. 22: DRY CONTACT WIRING DIAGRAM

Step 3. Go to the product service manuals (Chapter 6) for

instructions on how to map to the digital inputs.

Output Relay Wiring

The HI 3000 Series has 4, OPTO22 solid state relays. Please
see Relays in Chapter 2 - Specifications of each product Ser-
vice manual for details.

Step 1. Wire into the output relay of your choice.
Step 2. Go to the product service manuals (Chapter 6) for

instructions on how to map to the output relays.

Load Point Connections

FIG. 23: REAR PANEL/LOAD POINT CONNEC-

TIONS

C2

®

Load Point Connection

WARNING: L

OAD

CELL

CABLE

LENGTH

HAS

BEEN

CAL

-

CULATED

INTO

C2

CALIBRATION

DATA

. H

ARDY

P

ROCESS

S

OLUTIONS

RECOMMENDS

THAT

YOU

DO

NOT

CUT

YOUR

A

DVANTAGE

OR

A

DVANTAGE

L

ITE

LOAD

SENSOR

CABLE

,

AS

YOUR

C2

ACCURACY

WILL

BE

AFFECTED

AND

THE

WARRANTY

WILL

BE

VOIDED

.

Cable color Code for C2 Load Points (left to right facing the
rear panel):

Shield

Ground Wire

C2-

Violet

C2+

Grey

EXC-

Black

SEN-

Brown

SIG-

White

SIG+

Green

SEN+

BLUE

EXC+

RED

Step 1. Remove the factory installed jumper from the ter-

minal block if you are connecting an 8 wire cable
from the junction box.

Step 2. Connect the cable (Recommended load cell cable:

Hardy Instruments Prt. # 6020-0001) wires to the
Channel 1 terminal block according to the cable
color chart.

NOTE:

To purchase Hardy Load Cell cable, contact your
local Hardy Representative or Distributor.

Step 3. Plug the terminal block into the Channel connector

on the rear panel.

Step 4. For more information concerning C2 Load Point

connections, consult the individual HI 3000 Series
Product, Installation and Service Manual.

Non-C2 Load Cell Connection

Cable color Code using Hardy cable from a junction to an
instrument - for Non-C2 load cells:

Shield

Ground Wire

C2-

Not Used

C2+

Not Used

EXC-

Black

SEN-

Brown

SIG-

White

SIG+

Green

SEN+

Blue

EXC+

Red

Step 1. Remove the factory installed jumper from the ter-

minal block if you have 6 wire load cell cable that
includes sense wires from the load cell or junction
box.

Step 2. Connect the cable (Recommended load cell cable:

Hardy Prt. # 6020-0001) wires to the J9 terminal
block according to the Non-C2 cable color chart.

Step 3. Plug the terminal block into the Channel 1 (J9) con-

nector on the rear panel.

Step 4. For more information concerning Non-C2 Load

Point connections, consult the HI 3000 Series
Installation and Service Manual.

RS 232 Connection

The RS 232 Connection provides for a serial port for trans-
mission to a Printer or a Scoreboard. To configure the Printer
or Scoreboard see Chapter 4/Serial Port Parameters in the HI
3030 Service Manual.

Gnd

1 2 3 4

5