HI-3000 Series

80

Operation and Installation

appears in the Current Mappings text field. (See
Fig. 265)

FIG. 265: HARDY CONTROL-LINK FLOAT OUT/

WORD 0 (EFO0) SET AS MAPPING DESTINATION

Step 7. Click on the “Jump to Source Page” button. The

Mapping Setup Page 2 Appears. (See Fig. 266)

FIG. 266: MAPPING SETUP 2/SOURCE SELECTION

PAGE

Step 8. Click on the Process Data Pull Down List. (See Fig.

267)

FIG. 267: MAPPING SETUP 2/SELECTING PROCESS

DATA/GROSS WEIGHT

Step 9. Click on Gross Weight. (See Fig. 267) Gross Wt

appears in the Process Data field with a Chan field
containing the number 1 which is the channel we
want.

FIG. 268: PROCESS DATA/GROSS WEIGHT

CHANNEL 1 (HF14)

Step 10. Click on the Select button. “HF14” appears to the

right of the assignment equals sign.

FIG. 269: GROSS WEIGHT-CHANNEL 1 ASSIGNED

TO HARDY CONTROL-LINK FLOAT OUT

Step 11. Click on the Map button to map the assignment

statement. (See Fig. 269)

FIG. 270: GROSS WEIGHT ASSIGNED TO HARDY

CONTROL-LINK FLOAT OUT

Step 12. Click on the left arrow to return to Page 1.
Step 13. Click on the Network pull down list and select

Hardy Control-Link Float Out again.

Step 14. This time type the number “2” in the Word field.

(See Fig. 271)

NOTE:

Remember the Items in the Host Message on the
OPC Server consist of two words.

FIG. 271: HARDY CONTROL-LINK FLOAT OUT/

WORD 2

Step 15. Click on the Select button. Hardy Control-Link

Float Out Word 2 (EFO2) appears in the Current
Mappings field. (See Fig. 272)