Advanced setting mode – KEYENCE IL Series User Manual

Page 3

Advertising
background image

3

04 Alarm setting

Set the Alarm setting from [Default], [Clamp] or [User setting].
[Default]: When the measurement fails (out of the measurement range)
for 6(sampling rate set to 0.33ms) or 7(sampling rate set to except

0.33ms) times, the previous value is held until the next measurement is
successful.
[Clamp]: When the measurement fails (out of the measurement range),

the previous value is held until the next measurement is successful.
[User setting]: When the measurement fails (out of the measurement
range) for a certain number of times, the previous value is held until the

next measurement is successful.

05 Output state

Set the output status from [Normally Open] or [Normally Closed].

Advanced setting mode

How to enter the Advanced setting mode

1. Press the [MODE] button for at least 2 seconds.
2. Press the [MODE] or left/right arrow

button until the sub display will show
“End”.



3. Press the up/down arrow button. The

main display will show “Pro”.



4. Press the [MODE] or left/right arrow button and select the function

number.

Function

Explanation

06 Hold function

Set how to hold the value from

[Sample], [Peak], [Bottom], [Peak to
peak], [Auto peak] and [Bottom
peak].

07 Timing input

Set how the timing input for the hold

function is used from [Level] and
[Edge].

08 Delay timer

Set the delay timer for the judgment
output from [Off], [ON-delay],

[OFF-delay] and [1-shot].

09 Hysteresis

Set the hysteresis for the judgment

value.

10 Analog output scaling

Set how the analog voltage or
current should output, related to the

value.

11 External input

Assign each input function to the 4

external inputs.

12 Bank switching method

Set the way how to switch banks
from [Button] or [External input]

13 Memorizing the zero shift

Set if the shift target value has to be

memorized to the EEPROM or not,
when the zero shift was input.

14 Interference prevention

Set to use the interference
prevention function.

15 Number of digit

Set the number of digits from

[Default], [0.001], [0.01], [0.1] and
[1].

16 Power saving

Set to use the power saving
function.

17 Head display mode

Set the lighting conditions for the

light on the sensor head.

18 Display color

Set the color of the judgment light,
and the color of the main display.

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

GO

HOLD

CALC

CHECK

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

GO

HOLD

CALC

CHECK

1114-1

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

GO

HOLD

CALC

CHECK

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

GO

HOLD

CALC

CHECK

Advertising