9. Calibration

9.1 Calibration Items

You Ciin conduct Ihc lollowing in Uic; calibration mode:

• Checks or calibrates the zero and spar of the LF404 converter by using a simulation signal,

• Checks of Ihe excitation current.

To

change the mode to the calibration mode> see 7.3,1, “Mode Gtange,”

IMPORTANT

To check or change the zero and span of the converter, follow the procedure described

below. However, these are already checked and calibrated when shipped from the

factory. Do not change these settings unless it is necessary to calibrate in the field.

Calibration items are listed below. See each section for detailed procedure.

Section

Configuration item

Display example

Page

9.2.1

f)

%

flow rale calibration

NlîFLOWCAL 0%

0.0

%

102

9,2.2

50

%

flow rate calibration

N2:FLOW SIG.50%

50.0

%

103

9.2.3

1

ÜÜ

%

flow rate calibration

N3;FLOW CAL10Ü%

100.0

%

104

9.2.4

Excitation current

N4:EX. CURR.DSR

0.2100 A

105

-