10<3 Multi-range Functions

Multi-rftngc funclions can be set under the configuration item “RANGE TYPti.'" Refer to

8

.

2

,

6

, “Span (Range)/’ Four types of multi-range configurations are avaiJahIc as shown

below:

(1) Automatic selection of unJdirectional flow multi-range

(2) Automa tic se lect ion of bi d ireci ional fio ws multi ■ range

(3) Remote selection of unidirectional flow multi-range with an external signal

(4) Remote selection of bidirectional flows multi-range with ал external signal

Proceed as follows to use the multi-range functions.

Range setting

Set as follows referring to 8.2.5, “Span (Range)/’

1.

Select ^^RANGE TYPE."

2

. Set the span for ranges J lo 4,

3.

Sei the hysteresis value.

DO/DI function setting

Set DOl and/or D02 (option) to use them as range outputs. Refer to 8.2.9, “Digital

I/O"

To select ranges with a rernote signal, use DT (option) as a remote signai. Refer to

8.2.9, “Digital FO"

-

1 1 3

-