1

) Cle:ming, sLeriliziilion and drying

According lo ihe rules of cleaning (included sterilization and drying) pipe line for sysiem

and the specifications described m chapter. 15, clean (included sterilize and dry) the pipe

wall inside of detector and the gaskets.

Gaskets arc the special one for J.F494. In case of installing new gaskets, see Tablel5.3

■*Type Specification Code Maintenance Parts’’ and order it to your nearest Toshiba

representative.

(2) Mainienance for usual and stable flow measurment

if the fluid to he measured contains slurry; a high concentration of electrically conductive

solids^ the slurry may accumulate as sticking on the pipe wall inside of detector.

The Slicking causes a reduction of flow measuring outputs when it is not cleaned inside

pipe of detector for

a long time,

The flowmeter needs to make sure it remains around the wall inside pipe or not when the

following condition had happened.

[ 1 ]Thc flowmeter becomes a reduction of its output.

[

2

] i he flowmeter can not improve its output in spite of being calibrated*

'rhe flowmeter is required to remove along wnth clean it using a soft brush. It will be

come buck the usual and stable flow measurement after cleaning.

'I he pipe wall is never scratched when cleaning.

^Toshiba recommends to clean inside pipe periodically when this condition will be­

come easily.

*lt is belter Lo choose the suitable diameter which becomes more than 3 m7S as its

flow velocity to avoid this condition generally.

Notes: Before uninstalling the ilowmeier(deLector) from the pipe linc^ necessarily confirm

Ihe fluid of temperature and chemical property, and empty the fluid from detector.

Prepare the way to protect operator and system from bad influence; high temperature

fluid and chemical property

-