Important, I if connection pipe is conductive, If connection pipe is non-conductive – Toshiba LF494 User Manual

Page 27: Figure 5.2 grounding procedure, Toshiba

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

TOSHIBA

6 F 8 A 0 7 7 4

IMPORTANT

(

1

) The grounding terminal of the LF494/UH04 flo^mEtcr should be grounded with

1 0 0

ohm or less ground resiiilance. Use a heavy copper braid or wire

(cross-seciioniil area

5 . 5

mn:" minimum) to ground the terminal and make it as

short as possible. The icnninal is M4 size and an M4-sizc crimped ring lug should

be used lo connect the wire to the tenninaU Avoid a common ground where earth

current may flow . An independent ground is preferable^ See Figure

5 . 2 .

for a

conductive pipeline grounding and noii-conducrive pipeline grounding procedures.

(2) To prevent a wo-poinl grounding, ground the shielded cable on the receiving

inshiimenl side.

I

If connection pipe is conductive:

Connect between the grounding terminal
and both ends of the mating flanges with

a heavy copper braid or wire

(cross-section 5.5 min‘ minimum).

If connection pipe is non-conductive:

Use a heavy copper braid or wire

(cross-scciiiin 5.5 mm^ minimum) to

ground he terminal with UK) ohm or

less ground resistance*

Figure 5.2 Grounding Procedure

-