13Л Maintenance

Ш

Calibration

The LR04 converter has a reference signal generating circail. This reference signal can he

used to check the zero and span of the converter for the purpose of instrumentation

maintenance or periodical inspection. Sec Chapter 9» “Calibration,'’

fuse

The fuse can be taken out by unscrewing the cap of the fuse holder. Check that the fuse is

not damaged. The fuse has to be replaced periodically. The recommended replacement

period is 3 years.

Glass tube fuse (normal blow type) 1 piece

1 A, 230 V for 100 to 240 V ac or 110 V dc power supply

2A, 25(] V for 24 V dc power supply

Dimensions:

Diameter 5.2 mm x: 20 mm

Type of flise used:

Rating:

LCD display (optional)

If the characters displayed on the LCD are dimmed or blurred^ the LCD display should be

replaced. To extend the life of the flowmeter, replace the l.CD early. To check and replace

the LCD display, contact your nearest Toshiba représenta live.

Power supply unit

Llectronic components deteriorate faster when the ambient temperature is high. The life of

the power supply unit in the converter is 9 to 10 years if the ambient temperature is 40"

C y

and 3 to

6

years if it is 50" C. To extend the life of the flowmeter, we recommend you

replace the power supply unil early. Contact your nearest Toshiba representative for a

flowmeter inspection or unit replacement.

- -