Intek RheoVac DR User Manual

Page 15

Advertising
background image

- 13 -

3.5

COMMUNICATIONS

A typical RheoVac DR instrument has an Ethernet port and two serial communications choices (RS-232
or RS-422). The ethernet port can communicate via Windows

®

network Modbus or OPC. Eight 4-

20mA output signals for the probe pair are optional. Each process variable is a linear, fully temperature
and pressure compensated value on any of these readable outputs. If supplied, all 4-20 mA output
signals are scaled such that 4 mA represents 0% of the rated full scale value (except temperature, which
is 0°C) and 20 mA represents 100% of the rated full scale value (temperature is 100°C). The standard
full scale values and definitions of all process variables are listed in Table III. Communication with the
RheoVac unit using Intek supplied software is discussed in Appendix A and Appendix B of this manual.
Note: Intek recommends using the network connection for all data transmissions and RheoVac
communications.

Table III. Process Variable Definitions and 4-20 mA (optional) Range

PROCESS VARIABLE

4-20 mA

FULL

SCALE

PROCESS VARIABLE DEFINITION

ACTUAL VOLUME FLOW

ACFM [m

3

/hr]

5000

[8,500]

The actual volumetric flow rate of gases leaving the vacuum dryer. It is a
measure of exhauster capacity. Decreased capacity means pump
degradation.

TOTAL MASS FLOW

lbs/hr [kg/hr]

10000

[4,536]

The total mass flow rate of the flowing gas. Note: this value is not a
measure of air in-leak. It is a measure of total vapor and air removal and
exhauster operating capacity.

WATER VAPOR MASS FLOW

lbs/hr [kg/hr]

8000

[3,629]

The water vapor component of the flowing gas being removed from the
vacuum dryer.

RheoVac PRESSURE

"Hg [mm Hg]

30

[762]

Absolute pressure at the RheoVac DR probe head. Should be equal to or
less than turbine back pressure.

WAT ER VAPO R SPE CIFIC

VOLUME

(ft³/lb)

N/A

The inverse density of the water vapor present in the line. Not supplied as
4-20 mA output.

WATE R to AIR MASS

RAT IO

20

Ratio of water vapor flow rate to dry air flow rate. Defines “vacuum
quality.”

RELATIVE SATURATION

[%]

100

The percent concentration of water vapor in the extraction line relative to
saturation.

PARTIAL PRESSURE,

WATER

“Hg (mm Hg)

N/A

The partial pressure of water vapor in the vaccuum line. Not supplied as
4-20 mA output.

AIR MASS FLOW

lbs/hr [kg/hr]

1000

[453.6]

Actual measure of air mass flow rate passing the RheoVac DR sensor head,
normalized to standard conditions (70°F, 29.9" HgA).

RheoVac TEMPERAT URE

°F [°C]

212

[100]

4 mA = 0°C/32°F 20 mA = 100°C/212°F
Temperature of the flow media at the RheoVac DR probe head.