Operating conditions, Notice – VACUUBRAND DCP 3000 User Manual

Page 16

Advertising
background image

page 16 of 64

• Adopt suitable measures in case of differences from

recommended conditions, e.g., using the equipment

outdoors, or in an atmosphere of conductive pollution

or external condensation.

• Do not operate this product near flames.

To the best of our knowledge the equipment is in com-

pliance with the requirements of the applicable EC-direc-

tives and harmonized standards (see ”Declaration of Con-

formity”) with regard to design, type and model. Directive

EN 61010-1 gives in detail the conditions under which the

equipment can be operated safely (see also IP degree of

protection, “Technical data”, pg. 21).

Operating conditions

➨

This device is not approved for operation in potentially

explosive atmospheres.

Do not operate the device in

potentially explosive atmospheres.

➨

Devices without the

”

`

”

mark on the rating plate

are

not approved for operation with dangerous or

explosive gases or with potentially explosive or in-

flammable substances. Do not operate the device

with dangerous or explosive gases or with poten-

tially explosive or inflammable substances.

➨

Devices bearing the

”

`

”

mark on their rating plates

are

approved for operation with potentially explo-

sive atmospheres according to their classification

II 3G IIC T3 X according to ATEX, but they are not

approved for operation in potentially explosive at-

mospheres (see section “

`

Important information:

Equipment marking (ATEX)”, pg. 19)

.

• Ensure that the materials of the vacuum gauge’s wet-

ted parts are compatible with the substances in the

vacuum system, see section “Technical data”, pg.

21.