Cleaning the pressure transducer, Notice – VACUUBRAND DCP 3000 User Manual

Page 52

Advertising
background image

page 52 of 64

The vacuum gauge itself is maintenance free.
Condensate and deposits will affect the measurement results and/or the

adjustment of the pressure transducer.

Attention: Never use a pointed or sharp-edged tool to

clean the pressure transducer.

VSK 3000:

Never touch the ceramic diaphragm of the pressure trans-

ducer with hard objects.

VSP 3000:

The interior of the pressure transducer is highly sensitive!

Do not insert fingers or tools into the measuring chamber.

➨

Fill the measurement chamber with a solvent (e.g., benzene) and al-

low sufficient cleaning time. Observe all regulations concerning usage

and disposal of solvents!

➨

Drain the solvent and dispose of in accordance with regulations. Re-

peat cleaning if necessary.

➨

Rinse the measurement chamber several times with alcohol in order

to remove all solvent residues.

➨

Allow the pressure transducer to dry.

➨

Readjust the pressure transducer if necessary, see sections “Read-

justment of VSK 3000”, pg. 54 / “Readjustment of VSP 3000”, pg.

56.

NOTICE

Cleaning the pressure transducer