Adjustment under vacuum

An adjustment under vacuum is only possible

in the lower pressure range.

➨

Evacuate the measurement connection of the VSP 3000 pressure

transducer preferably below 7.5x10

-4

Torr (1x10

-3

mbar), e.g., by apply-

ing a high vacuum pump. Allow the pressure transducer a

minimum

of 20 minutes to preheat at this pressure.

➨

In ”Configuration” menu, select program ”Adjustment” at the DCP 3000.

+

The reading is automatically adjusted to ”zero”. This value is fixed and

cannot be changed.

➨

Press the selection knob to confirm.

0 Torr

➨

In ”Configuration” menu, select program ”Adjustment” at the DCP 3000.

➨

The reading is automatically adjusted to ”750 Torr” (”1000 mbar”). This

value is fixed and cannot be changed.

➨

Press the selection knob to confirm.