General information on handling the VSP 3000

The VSP 3000 pressure transducer which has been developed particu-

larly for use in chemical laboratories is a pressure sensor working in the

fine vacuum range. Over the whole measuring range the pressure read-

ing is indicated in exponential notation.

Allow the VSP 3000 a period of approx. 20 minutes to preheat and to

meet its specifications. Even in case of the vacuum gauge being switched

off, the VSP 3000 stays ready for operation. To de-energize the pressure

transducer, unplug either its VACUU•BUS line or the power supply of the

vacuum gauge.

The interior of the pressure transducer is highly sensitive!

Do not insert fingers or tools into the measuring chamber.

VACUU•BUS

Read the pressure transducer with the DCP 3000 vacuum gauge, con-

nected via VACUU•BUS line. Maximum cable length inside buildings:

30m (~32yd) (extension cable VACUU•BUS 2m (6.6ft): order-no. 612552

/ extension cable VACUU•BUS 10m (32.8ft): order-no. 2618493).

Do not use more than one DCP 3000 within the same VACUU•BUS

system. Several DCP 3000 in the same VACUU•BUS system will inter-

fere with each other and result in error messages of the connected com-

ponents (pressure transducers, valves).