Safety during operation

➨

Adopt suitable measures to prevent the release of

dangerous, toxic, explosive, corrosive, noxious or pol-

luting fluids, vapors and gases.

+

Never operate this vacuum gauge if it has a damaged

cord or plug. If the vacuum gauge is not working prop-

erly, has been dropped or has fallen into water, contact

your service provider.

+

Prevent any part of the human body from coming into

contact with vacuum.

Operation with pressure transducer

VSK 3000:

+

Attention: At pressures above approximately 795 Torr

(1060 mbar) the pressure reading becomes incorrect

due to saturation of the pressure transducer. The dis-

play flashes. Release pressure immediately!

Risk of

bursting!

Operation with pressure transducer

VSP 3000:

+

Attention: Maximum pressure reading: 7.5*10

2

Torr

(1*10

3

mbar). Pressure values above 7.5*10

2

Torr

(1*10

3

mbar) can not be displayed.

Danger of unno-

ticed overpressure! Risk of bursting!

Operation with pressure transducer

MPT 100:

+

See instructions for use of MPT 100.

Electronic equipment is never 100% fail-safe. This may

lead to an ill-defined status of the equipment or of other

connected devices. Provide appropriate protective mea-

sures to allow for the possibility of failure and malfunction.

The protective measures must also allow for the require-

ments of the respective application.