Read commands – VACUUBRAND DCP 3000 User Manual

Page 46

Advertising
background image

page 46 of 64

Command

Operation

Response

Description

IN_PV_1

current pressure

XXXX mbar/Torr/

hPa*

X.XXE

±X mbar/

Torr/hPa*

unit according to preselections

IN_PV_Sx

current pressure

of pressure trans-

ducer x

XXXX mbar/hPa/

Torr*

X.XXE

±X mbar/

Torr/hPa*

pressure of pressure transducer x
(order of numbering according to dis-
play in ”Sensors” menu)

IN_PV_X

current pressure

values of all con-

nected pressure

transducers

XXXX.X XXXX.X

...mbar/Torr/hPa*

X.XXE±X X.XXE±X

...mbar/Torr/hPa*

pressure values of all connected pres-
sure transducers

IN_PV_D

data readout

XXX.XX:XX:XX X

...mbar/Torr/hPa*

read out data from DCP 3000 memory

IN_CFG

device set

preselections

yXXXXXX

X0XXXXX
X1XXXXX
X2XXXXX
XX0XXXX
XX1XXXX
XXX0XXX
XXX1XXX
XXXX0XX
XXXX1XX

XXXXXyX

XXXXXXy

y: 0......D: language

**

(hexadecimal)

pressure unit mbar
pressure unit Torr
pressure unit hPa
acoustic signal off
acoustic signal on
venting valve not connected
venting valve connected
fault indicator not connected
fault indicator connected
y: 1......8: number*** of active pressure
transducer
y: 1......8: pressure transducer quantity

IN_ERR

fault status

0XXXX
1XXXX
X0XXX
X1XXX
XX0XX
XX1XX
XXX0X
XXX1X
XXXX0
XXXX1

no fault at venting valve
fault at venting valve
no overpressure
overpressure
no fault at pressure transducer
fault at pressure transducer
no external fault
external fault
last interface command correct
last interface command incorrect

IN_VER

version

DCP 3000 Vx.xx

software version

IN_SP_1

time interval for

sending

XX:XX

mm:ss (minutes:seconds);
time interval for sending

IN_SP_2

time interval for

recording

XX:XX

mm:ss: (minutes:seconds);
time interval for recording

Read commands