Interface parameters, Setting of the interface – VACUUBRAND DCP 3000 User Manual

Page 43

Advertising
background image

page 43 of 64

The vacuum gauge DCP 3000 is equipped with a serial interface (RS

232C, nine-pin Sub-D-plug).
+

Plug-in or remove the cable (cable RS 232C) from the interface only if

the equipment is switched off.

+

The interface is

not electrically isolated from the measuring circuit.

+

For optimal electromagnetic compatibility assemble an interface filter

(cat. no.: 638235).

+

The serial interface is prepared for the connection of a commercial

Bluetooth adapter for wireless communication (current supply for Blue-

tooth adapter on pin 9, see ”Pin assignment”).

+

Settings via interface are stored.

Interface parameters

Setting of the interface

Set the interface parameters directly at the DCP 3000. The factory set

values are underlined.

Edit and confirm the interface parameters in the ”Configuration” menu in

”RS-232” submenu using the selection knob.

Baud: 2400, 4800, 9600 or 19200

Parity: 8-N-1, 7-O-1 or 7-E-1

Handshake: Off, Xon-Xoff or RTS-CTS

Send: Off or 1......3600 s

Data: Send

A maximum of twenty commands per second is possible.

The commands have to be written in capital letters.

Command and parameter have to be separated by a blank.

The string is terminated with <CR> or <LF> or <CR><LF>.

The response of the controller is always terminated with <CR><LF>.

Numerical values and parameters can be written without leading zeros.

The response of the controller always includes leading zeros.