VACUUBRAND DCP 3000 User Manual

Page 36

Advertising
background image

page 36 of 64

To switch off automatic sending select ”off” or stop via serial interface.

Send recorded data. See section “Notes concerning the data logger

function (recording and sending of data)”, pg. 44.

+

Sensors: Selection of the pressure transducer to be controlled.

Maximum number of pressure transducers of the same type (VSK

3000 or VSP 3000 or VSK 3000 configured as reference) connected

to one DCP 3000: four.

You may replace up to four VSP 3000 pressure transducers by MPT 100

pressure transducers.

+

Display: Selection of the device parameters ”Brightness” between 0

- 100%; ”Contrast” between 0 - 100%; ”Sound” ”On” or ”Off”; ”Units

”mbar”, ”hPa” or ”Torr”; ”Language” ”German”, ”English”, ”French”,

”Italian”, ”Spanish”, ”Turkish”, ”Korean”, ”Chinese”, ”Portuguese”,

”Russian”, ”Polish”, ”Dutch”, ”Japanese”, ”Finnish”.

+

Setpoint: Maximum four “On” setpoints and four ”Off” setpoints can

be preset after appropriate configuration of valves and / or digital I/O

interfaces (see section “Setpoints”, pg. 41).

+

Log data: Recording and internal storage of time and pressure data

of up to 4 pressure transducers (Log Data ”Off” or interval 1 . . . 3600

seconds)

The screen-shot shows the factory-set values.

Configuration
Adjustment 760 Torr

RS-232 . . .

Sensors . . .

Display . . .

Setpoint . . .

Log Data Off

- - - - - - - Graphic - - - - - - -

- - - - - - - - Back - - - - - - - -

Advertising