Electrical requirements, Service line requirements – Labconco Purifie Horizontal Clean Benches 36125 Series User Manual

Page 13

Advertising
background image

9

Product Service 1-800-522-7658

Chapter 2: Prerequisites

Electrical Requirements


The different Purifier Clean Bench models have the
following electrical requirements:

Model #

Requirements

3600000 through 3600009

115 VAC, 60 Hz, 12 Amps

3600020 through 3600029

230 VAC, 50 Hz, 7 Amps

3612500 through 3612509

115 VAC, 60 Hz, 12 Amps

3612520 through 3612529

230 VAC, 50 Hz, 7 Amps

3610000 through 3610009

115 VAC, 60 Hz, 16 Amps

3610020 through 3610029

230 VAC, 50 Hz, 8 Amps


All Purifier Clean Benches with model numbers ending
in 00 through 09 are designed for operation at 115
volts, 60 Hz, alternating current. Clean Benches with
model numbers ending in 20 through 29 are designed
for operation at 230 +/- 20 volts, 50 Hz alternating
current. A dedicated outlet with a circuit breaker rated
at 20 amps should be located as close as possible to the
right rear side of the bench, at a height even with, or
higher than, the bottom of the bench in its final
location. If your electrical outlet is distant from this
location, contact Labconco’s Product Service
Department for information on longer cords.

Service Line Requirements


All service lines to the Clean Bench should be quarter
inch outside diameter, metal, and equipped with an
easily accessible shut-off valve should disconnection be
required. If the service line pressure exceeds 40 PSI, it
must be equipped with a pressure regulator to reduce
the line pressure.

Table 2-1

Advertising