Model Series

36000

36125

36100

Width (feet)

3

4

6

HEPA Filter Dims. (in.)

36 x 30 x 6

48 x 30 x 6

72 x 30 x 6

Average Velocity Reading Grid (inches)

12 x 12

12 x 12

12 x 12

Grid Distance from Diffuser (inches)

6

6

6

Grid Distance from Sidewalls (inches)

6

6

6

Average Velocity (FPM)

90 ±10

90 ±10

90 ±10

Work Area (ft

2

) 7.5

10

15

Exhaust Volume Range (CFM)

600-750

800-1,000

1,200-1,500

Number of Laskin Nozzles needed

1

2

2

APPROXIMATE SETTINGS FOR NEW FILTERS

Mag Gauge Reading

0.7 – 1.0

0.7 – 1.0

0.7 – 1.0

Motor HP

1/3

1/2

2 1/3

Speed Winding used

High

High

High

Motor Capacitator

15

15

15

Fluorescent Light(s)

F25T8/TL741 F32T8/TL741

F72T12CW

UV Light(s) (Option)

G30T8

G30T8

G64T5L/CB