Chapter 6: Maintaining Your Clean Bench


Weekly

• Wipe down the interior surfaces of the clean

bench with a mild household detergent or
cleaner and allow to dry.

• Using a damp cloth, clean the exterior surfaces

of the clean bench, particularly the front and top
of the clean bench to remove any accumulated
dust.


Monthly (or more often as required)

• Check all service valves, if so equipped, for

proper operation.

• Check the prefilters and replace if necessary.

The prefilters should be replaced at least
quarterly.

• All weekly activities.


Quarterly

• Replace the prefilter. See page 57 for ordering

information.

• All monthly activities.


Annually

• Have the clean bench recertified by a qualified

certification technician.

• All quarterly activities.


Biannually

• Replace the fluorescent lamp if required.

• Replace the UV lamp (if included).

• All annual activities.

Service Operations


Resetting the Circuit Breaker:

1. The circuit breaker is located next to the power cord

on the upper rear panel of the bench as shown in
Figure 6-1. If the circuit breaker trips, it can be
reset by pressing the white button in.