2. The lamp is located on the underside of the canopy,

between the fluorescent lamp and the diffuser.
Remove the UV lamp by grasping the plastic
endcap on the left end of the lamp. While
supporting the center of the lamp, push the endcap
towards the left.

3. Allow the right side of the lamp to drop out of its

socket, then pull the lamp straight down.

4. Make sure that the surface of the new lamp is clean

and free of dirt or fingerprints by cleaning its
surface with alcohol or glass cleaner.

5. Install the new lamp by reversing the above steps.

The service operations listed in the
rest of this chapter should only be
performed by a qualified certifier.


Speed Control Adjustment:

Adjusting the speed control will have
an effect on the air velocities and the
effectiveness of the clean bench. Only
a qualified certification technician, as
part of the recertification process,
should adjust the speed control.