Labconco Purifie Horizontal Clean Benches 36125 Series User Manual

Page 52

Advertising
background image

48

Product Service 1-800-522-7658

Chapter 6: Maintaining Your Clean Bench



4. After completing the speed control adjustment,

establish the average air velocity, using a calibrated
thermal anemometer with an accuracy of ± 3%.
Establishing a boundary of 6 inches from the sides,
top and bottom, measure the velocities in 6-inch
increments, 6 inches in front of the diffuser. This
methodology is more fully defined in the IEST
recommended practice #IEST-RP-CC0002.2

5. The average velocity should be 90 ± 10 FPM, with

all measured values falling within ± 20% of the
average.

6. Reinstall the prefilters.

Diffuser Removal:

1. Locate and remove the diffuser screws, shown in

Figure 6-3. Pull the diffuser straight out.