Connecting the clean bench to service lines – Labconco Purifie Horizontal Clean Benches 36125 Series User Manual

Page 23

Advertising
background image

19

Product Service 1-800-522-7658

Chapter 3: Getting Started


The clean bench is very top heavy!
Use caution when lifting or moving it.


4. Carefully lift the clean bench onto the top rails of

the base stand and slide it into position. The rear of
the bench should be flush with the rear of the base
stand and the sides of the clean bench will overhand
the side rails of the base stand approximately 0.25
inches on both sides.

5. Insert the 2.5 inch x 0.25 inch mounting bolts

through the hole in the underside of the base stand
and secure it with a flatwasher and lockwasher.

6. Repeat steps 5 and 6 for the remaining bolts.
7. Tighten all four bolts securely.

Connecting The Clean
Bench to Service Lines


The service lines (if any) should be connected to the
tube fitting(s) on the outside of the side wall as shown
in Figure 3-4. To install the tubing, follow these steps:

1. Ensure that the tubing is 0.25 inch outside diameter,

soft metal, and that the end has been deburred
completely. You will need a 0.25 inch tube fitting
with a 0.25 inch NPT male thread (not included) to
connect the tubing to the valve.

2. Make sure that the tube fitting nut on the fitting is

loose, but do not remove it. Look inside the fitting
to make sure the tube ferrule is there.

3. Push the tube into the fitting until it is properly

seated. The tube will go approximately 0.75 inch
into the fitting.

4. Tighten the tube fitting nut hand tight and then,

using a 7/16 inch wrench, tighten it at least three
quarters turn more.

5. Close the service valve in the clean bench and then

slowly open the shutoff valve on the service valve.
Inspect the fitting for leakage. Tighten the tube nut
slightly if needed.

!

Advertising