Chapter 2: Prerequisites

The use of flammable gases or solvents
should be avoided in the clean bench.
Open flames in the cabinet will
disrupt the laminar airflow in the
bench. If you feel that your procedure
requires the use of an open flame or
flammable materials, contact the
appropriate safety official.

The use of air or gases under high
pressure should be considered
carefully as they may seriously disrupt
the airflow patterns in the bench.

Space Requirements


The dimensions for the different models are shown in
Appendix B: Clean Bench Dimensions.

Overhead Clearance


In order for the clean bench to operate properly, there
should be at least six inches (150 mm) of clearance
from any overhead obstructions when the bench is in its
final operating position.

Your Next Step


After you have determined that the location you have
selected accommodates the installation and operational
requirements of your clean bench, you are ready to
begin installation. Proceed to Chapter 3: Getting
Started
.


!