Chapter 4: Performance Features and Safety Precautions


with it should perform Perchloric
acid procedures.

The hood is specially designed for
use with Perchloric acid. It should
not be used as a general-purpose
laboratory fume hood. Because of
extreme hazards, miscellaneous
work should not be performed in
this hood.

Safety requires regular washdown
of fume hood interior, ductwork,
and the movement of air
preferably after each experiment.
The baffle on this hood may be
removed for access to all surfaces
as a follow-up to washdown
procedures. All deposits should be
flushed away. Utilize lowest
quantities of Perchloric acid to fit
procedural requirements.

All apparatus used within the hood
interior should have inorganic
coatings and lubricants.

Although, the Protector Perchloric
Acid Fume Hoods conform to all
generally accepted standards of
design for this type of equipment,
we accept no responsibility or
liability for accidents that may
possibly occur in the use of
Perchloric acid in our hoods.

Radioisotope materials are not
recommended for use in this hood.
Only use Perchloric acid.

AVOID CROSS DRAFTS AND
LIMIT TRAFFIC IN FRONT OF
THE HOOD. AIR
DISTURBANCES CREATED

)

!