Chapter 7: Modifying Your Protector Perchloric Acid Hood

Installing Additional Service
Fixtures

Additional service fixtures can be installed. The fume hood
is factory set to accept up to four valves per side, but the
liner is not drilled for extra fixtures. Contact Labconco
Customer Service for information.








Figure 7-2 Knob

Figure 7-3 Valve Figure 7-4 Stainless Connector

Installing Guardian™ Digital
Airflow Monitor or Guardian™
Jr. Airflow Monitor

The Guardian Digital Airflow Monitor P/N 9743211
continuously monitors face velocity through the fume hood
opening. The Guardian Jr. Airflow Monitor P/N 9743202
continuously monitors airflow through the exhaust. The fume
hood right corner post is factory prepared to mount either
monitor. To mount the Guardian Digital Monitor the hood
liner needs a 0.50" diameter hole mounted on the right side; 2"
above the sash handle at its uppermost height and 6" behind
the sash. Contact Labconco Customer Service to order.